Spectrometer ModelBrewer MK IV
F/number6
Diffraction grating
Number1
Typeplane
holographic
Lines per millimeter1200
Blaze [nm]
Diffraction orderthird
Dispersion1nm/mm
PMT9789QA
Bandwidth [nm]0.6
Step [nm]
usual0.5
finest0.1
range [nm]286 to 363
Diffuser materialTeflon
Weatherproof?Yes
Automatic?Yes
Temperature
Stabilized optics?No
Stabilized detector?No
Dark current removed?Yes
Stray light removed?Yes
Wavelength regist. [nm]302.3
Primary lamp [W]1000
Secondary lamp [W]50